Collage Site uses a calibration scan to match the color of the Ricoh Theta Z1 image to a point in the CR-P1 scan.
Collage Site performs color scans using parameters calculated from system calibration (initial calibration and subsequent calibration updates).
Calibration scan images are transferred to Collage Site at the same time as they are scanned and are ready for use; to colorize a PanoCam scan, you must capture at least one initial PanoCam To colorize a PanoCam scan, at least one first PanoCam calibration scan must be captured with the same Ricoh Theta Z1 camera used for the scan project. Calibration data is shared across all projects for one user.
For best results:
- The CR-P1 scanner and Ricoh Theta Z1 camera should be turned on approximately 15 minutes before calibration scanning to reach operating temperature.
- Calibration scans should be performed in an indoor environment such as an office or room.
- Walls should be a minimum of 2 meters (6 feet) and a maximum of 5 meters (16 feet) away from the scanner.
- Objects and patterns should be distributed on the floor, walls, and ceiling.
- Avoid calibrating in rooms with bright white walls, dark environments, open doors, people moving around, ceiling fans, or other moving objects.
- Make sure there are no mirrors, glass walls, or other reflective surfaces in the scanner's field of view.
- The scanner/tripod set should be placed on a flat, solid floor.
If the calibration scan does not produce the proper color overlap, perform the calibration scan again. Calibration scan should be performed again in the following cases:
- When using a camera or scanner other than a previously fully calibrated pair
- Before scanning a project that contains many scans
- If the scan lacks quality in color
- Scanning in environments with extreme temperature fluctuations or low/high temperatures
- If the mounting bracket moves or bumps against
- If the color overlap result indicates an error
The calibration process creates a 1/5 / 2x image quality scan that includes two PanoCam, color capture of the internal camera with equally weighted metering. The entire system calibration process takes approximately 3.5 minutes.